10-10
WARNING
No alarm is provided for the temperature measurement.
CAUTION
Used probe covers must be treated as infectious biological waste and disposed
of in accordance with current medical practices and local regulations.
10.3.2 Thermometer Buttons
Button Name Functions
Eject button Press the eject button to eject the probe cover from the probe.
°C/°F button When a temperature is in the display, you may press and hold
the °C/°F button to toggle between degrees Celsius and
degrees Fahrenheit.
Timer button Press and hold the timer button to enter the timer mode. Press
again to start the timer.
Scan button When the thermometer is on, press the scan button to initiate a
temperature measurement.
When the thermometer is off, press the scan button to turn on
the device.
10.3.3 Equivalence Mode Temperature
The thermometer can measure the ear temperature and acquire the equivalence
temperature.
A patient’s temperature at the site of mouth, rectum, or core would be slightly different
from the ear temperature. The thermometer compensates for the average difference in
temperature at each of these sites by adjusting the displayed temperature.
A checker/calibrator is available for this device. The device should be checked if it is
dropped or if it is stored at less than -25°C or above 55°C.
NOTE
Only the authorized personnel can adjust the data for Genius 2 thermometer
equivalence modes.