16-1 
16 Care and Maintenance 
 
 
The monitor should be maintained and cleaned on a regular basis. This chapter describes 
the basic cleaning, disinfection and test method. 
 
WARNING 
  Failure of the responsible individual, hospital or institution employing this 
equipment to implement a satisfactory maintenance schedule may cause 
undue equipment failure and possible health hazards. 
  The safety checks or maintenance involving any disassembly of the equipment 
should be performed by professional servicing personnel. Otherwise, undue 
equipment failure and possible health hazards could result. 
  If you discover a problem with any of the equipment, contact your service 
personnel or us. 
  The responsible hospital or institution shall carry out all cleaning and 
disinfection procedure specified in this chapter. 
  Do not open the equipment housings. All servicing and future upgrades must 
be carried out by the personnel trained and authorized by our company only. 
  No modification of this equipment is allowed. 
  The service personnel must be properly qualified and thoroughly familiar with 
the operation of the equipment. 
 
 
16.1 Cleaning and Disinfection 
In this chapter we only describe cleaning and disinfection of the main unit. For the 
cleaning and disinfection of other reusable accessories, refer to instructions for use of 
corresponding accessories. 
 
Keep you equipment and accessories clean. To avoid damage to the equipment, follow 
these rules: 
  Always dilute according the manufacturer’s instructions or use lowest possible 
concentration. 
  Do not immerse part of the equipment into liquid. 
  Do not pour liquid onto the equipment or accessories. 
  Do not allow liquid to enter the case. 
  Never use abrasive materials (such as steel wool or silver polish), or erosive 
cleaners (such as acetone or acetone-based cleaners).