7-4 
7.5.3 Setting SpO
2
 Sensitivity 
For Mindray SpO
2
 module, you can set [Sensitivity] to [High], [Med] or [Low] in the [SpO
2
 
Setup] menu; for Masimo SpO
2
 module, you can set [Sensitivity] to [Maximum] or 
[Normal] in the [SpO
2
 Setup] menu.   
 
When the sensitivity is set to [High] or [Maximum], the equipment is more sensitive to 
minor signals. When monitoring critically ill patients whose pulsations are very weak, it is 
strongly recommended that the sensitivity is set to [High] or [Maximum]. When 
monitoring neonatal or non-critically ill patients who tend to move a lot, noise or invalid 
signals may be caused. In this case, it is recommended that the sensitivity is set to [Low] 
or [Normal] so that the interference caused by motion can be filtered and therefore the 
measurement stability can be ensured. 
 
 
7.5.4 Changing Averaging Time 
The SpO
2
 value displayed on the monitor screen is the average of data collected within a 
specific time. The shorter the averaging time is, the quicker the equipment responds to 
changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging time 
is, the slower the equipment responds to changes in the patient’s oxygen saturation level, 
but the measurement accuracy will be improved. For critically ill patients, selecting 
shorter averaging time will help understanding the patient’s state. 
 
To set the averaging time: 
  For Mindray SpO
2
 module, select [Sensitivity] in the [SpO
2
 Setup] menu and 
then toggle between [High], [Med] and [Low], which respectively correspond to 
7 s, 9 s and 11 s. 
  For Masimo SpO
2
 module, select [Averaging] in the [SpO
2
 Setup] menu and then 
toggle between [2-4 s], [4-6 s], [8 s], [10 s], [12 s], [14 s] and [16 s]. 
 
 
7.5.5 Monitoring SpO
2
 and NIBP Simultaneously 
When monitoring SpO
2
 and NIBP on the same limb simultaneously, you can switch [NIBP 
Simultaneous] on in the [SpO
2
 Setup] menu to lock the SpO
2
 alarm status until the NIBP 
measurement ends. If you switch [NIBP Simultaneous] off, low perfusion caused by NIBP 
measurement may lead to inaccurate SpO
2
 readings and therefore cause false 
physiological alarms. 
 
 
7.5.6 Sat-Seconds Alarm Management 
With traditional alarm management, high and low alarm limits are set for monitoring 
oxygen saturation. During monitoring, as soon as an alarm limit is violated, an audible 
alarm immediately sounds. When the patient % SpO
2
 fluctuates near an alarm limit, the 
alarm sounds each time the limit is violated. Such frequent alarm can be distracting. 
Nellcor’s Sat-Seconds alarm management technique is used to reduce these nuisance 
alarms.