12-7
12.6 Managing Scorings
12.6.1 Importing the Scoring
You can import MEWS, NEWS and customized scorings into the monitor. Up to five scoring
protocols can be imported into the monitor.
1. Connect the USB drive to the USB connector on the monitor.
2. Select [Main] [Maintenance>>] [User Settings>>]Enter the required
password[Manage Scoring>>][Import Scoring>>].
3. In the [Import Scoring] menu, select the scorings to be imported. Then select
[Import]. A status message will report completion of the import.
NOTE
To load the customized scoring protocols to the USB drive, refer to Clinical
Scoring Config Instruction for Use (P/N: 046-007126-00).
12.6.2 Selecting Default Scoring
The monitor does not provide default scoring. To select a default scoring:
1. Select [Main] [Maintenance>>] [User Settings>>] Enter the required
password[Manage Scoring>>][Select Default Scoring>>].
2. Set the default scoring for the applicable patient category.
After the default scoring is set, when a patient category is changed, the monitor will
automatically use the default scoring.
12.6.3 Deleting the Scoring
To delete the scoring:
1. Select [Main][Maintenance>>][User Settings>>]Enter the required
password[Manage Scoring>>][Delete Scoring>>].
2. In the [Delete Scoring] menu, select the scoring to be deleted.
3. Select [Delete].
12.6.4 Loading a Scoring
The default scoring may not be appropriately for the new patient. You can load a scoring
so as that the scoring is appropriate for your patient.
To load a scoring:
1. Select [Scoring][Setup][Load Scoring>>].
2. Select the desired scoring to be loaded.
3. Select [Load Scoring].