A-4
AC power LED 1 (green)
Battery LED 1 (green)
A.5.4 Audio Indicator
Speaker
Give alarm tones (45 to 85 dB), key tones, pulse tone; power-on
self-check tone, support PITCH TONE and multi-level tone
modulation; alarm tones comply with IEC 60601-1-8.
A.5.5 Monitor Interface Specifications
Power 1 AC power input connector
Wired network 1 RJ45 connector
USB 2 standard connectors, USB 2.0
Equipotential Grounding
Terminal
1
Multi-Functions Connector 1 MINI-D RIBBON connector
A.5.6 Outputs
Nurse Call Signal
Amplitude
High level: >3V, providing a maximum of 3 mA output current
Low level: <0.5 V, receiving a maximum of 5 mA input current
Rising and falling time ≤ 1 ms
Alarm output (Network connector)
Alarm delay time from
monitor to remote
equipment
The alarm delay time form the monitor to remote equipment is
≤2 seconds, measured at the monitor’s signal output
connector.
A.5.7 Data Storage
Trends 5000 groups
A.5.8 Wireless Network
Standards MSD45N Wireless Module: IEEE 802.11a/b/g/n, support Wi-Fi
A.6 Measurement Specifications
A.6.1 SpO
2