5-2
5.2.1 Setting Default Configuration
The monitor will load the pre-set default configuration in the following cases.
The monitor is restarted after being switched off for above 120 seconds.
A patient is admitted.
A patient is discharged.
Patient category is changed.
The default configuration may come from the latest configuration, the factory default
configuration or the user configuration.
To set the default configuration:
1. Select [Select Default Config >>] in the [Manage Configuration] menu.
2. Select [Load Latest Config] or [Load Specified Config] in [Select Default Config]
menu.
When you select [Load Specified Config.], the configuration (adult, pediatric or neonate)
to be restored is subject to the patient category. This configuration can be either factory
configuration or saved user configuration. Take adult as an example, select [Default Adult
Config] and toggle between [Defaults] and user configuration(s).
NOTE
When the equipment starts, it shows what configuration is restored at the
prompt information area for about 30 seconds.
5.3 Saving Current Settings
Current settings can be saved as user configuration. Up to 3 user configurations can be
saved.
To save current settings:
1. Select [Save Current Settings As >>] in the [Manage Configuration] menu.
2. In the popup dialog box, enter the configuration name and then select [Ok].
5.4 Deleting a Configuration
1. Select [Delete Config >>] in the [Manage Configuration] menu. The popup menu
shows the existing user configurations on the monitor. Selecting [Config on USB
drive >>] will show the existing user configurations on the USB drive.
2. Select the user configurations you want to delete and then select [
Delete].
3. Select [
Ye s ] in the popup.