E.2 Device Enclosure and Accessories
E.2.1 Visual Inspection
Test Item Acceptance Criteria
The enclosure and
accessories
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
E.2.2 Contextual Inspection
Test Item Acceptance Criteria
The enclosure and
accessories
No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells, particularly
from ventilation holes).
No taped notes that may suggest device deficiencies or operator
concerns.
E.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
Main unit label
Integrated warning labels
E.4 Protective Earth Resistance
1. Plug the probes of the analyzer into the devices protective earth terminal and
protective earth terminal of the AC power cord.
2. Test the earth resistance with a current of 25 A.
3. Verify the resistance is less than limits.
LIMITS
ALL COUNTRIES R = 0.2  Maximum