9-4 
9.5 Measuring NIBP 
9.5.1 Preparing the Patient 
In order to minimize NIBP measurement errors, whenever possible check that the patient: 
  Is comfortably seated 
  Has legs uncrossed 
  Has feet flat on the floor 
  Has back and arm supported, and 
  The middle of the cuff at the level of the right atrium of the heart. 
 
NOTE 
  It is recommended that the patient relax as much as possible before the NIBP 
measurement is performed and that the patient does not talk during 
measurement. 
  It is recommended that the patient sit still for 5 min before the first 
measurement is taken. 
  The operator should not touch the cuff and tubing during the NIBP 
measurement. 
 
 
9.5.2 Preparation to Measure NIBP 
1  Power on the monitor. 
2.  Verify that the patient category is correct. If not, select the    hardkey → 
[Patient Demographics]→[Patient Category] and set the patient category to 
[Adult], [Pediatric] or [Neonatal]. 
3.  Connect the NIBP hose to the monitor.   
4.  Select the appropriate sized cuff by referring to the limb circumference marked on 
the cuff. 
  The width of the cuff should be 40% (50% for neonates) of the limb circumference, or 
2/3 of the upper arm’s length. The inflatable part of the cuff should be long enough 
to encircle at least 50% to 80% of the limb. 
5.  Apply the cuff to an upper arm or thigh of the patient and make sure the Φ marking 
on the cuff matches the artery location. Do not wrap the cuff too tightly around the 
limb. It may cause discoloration, and ischemia of the extremities. Make sure that the 
cuff edge falls within the marked range. If it does not, use a cuff that fits better. 
6.  Connect the cuff to the air tubing and make sure that the bladder inside the cover is 
not folded and twisted. Air must pass unrestricted through the tubing.