A-11
THP79JU Infrared Ear Thermometer
Technique Infrared
Measurement range 32°C to 42.2°C (89.6°F to 108°F)
Genius
TM
2 Tympanic Tethered Thermometer
Operating mode Adjusted mode
Response time 2 seconds
Measurement range
Ear: 33.0°C to 42.0°C (91.4°F to 107.6°F)
Oral: 33.6°C to 42.0°C (92.5°F to 107.6°F)
Core: 34.0°C to 42.0°C (93.2°F to 107.6°F)
Rectal: 34.2°C to 42.0°C (93.6°F to 107.6°F)
Calibrated accuracy limits
Ambient Temp Target Temp Accuracy
25°C (77°F)
36.7°C to 38.9°C
(98.1°F to 102°F)
±0.1°C (±0.2°F)
16°C to 33°C
(60.8°F to 91.4°F)
33°C to 42°C
(91.4°F to 107.6°F)
±0.2°C (±0.4°F)
Resolution 0.1°C or 0.1°F
A.6.5 CO
2
Standard Meet standard of ISO 80601-2-55
Technique Infrared absorption
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
CO
2
measurement range 0 to 20%
CO
2
accuracy
Full accuracy mode:
CO
2
concentration<1 %
1%≤ CO
2
concentration<5%:
5%≤ CO
2
concentration<7%:
7%≤ CO
2
concentration<12%:
12%≤ CO
2
concentration≤ 13%:
13% <CO
2
concentration≤ 20%:
CO
2
concentration>20%:
±0.1%
±0.2%
±0.3%
±0.4%
±0.5%
±0.43%+8%rel
unspecified
ISO accuracy mode:
Add ±0.3%ABS to the full accuracy
CO2 accuracy drift Meet the requirement for measurement accuracy within 6 hours.