B-4
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer or
the user of the device should assure that it is used in such an environment as described below.
Immunity test IEC60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conduced RF
IEC61000-4-6
3 Vrms
150 kHz to 80
MHz
3Vrms Portable and mobile RF communications
equipment should be used no closer to
any part of the system, including cables,
than the recommended separation
distance calculated from the equation
appropriate for the frequency of the
transmitter. Recommended separation
distances:
Pd 2.1
Radiated RF
IEC61000-4-3
3V/m
80MHz to
2.5GHz
3V/m Recommended separation distances:
80 MHz~800 MHz
Pd 2.1
800MHz-2.5GHz
Pd 3.2
Where,
P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d is the recommended separation
distance in meters (m).
b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
,should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion
band (2395.825MHz-2487.645MHz) is exempt from the ESSENTIAL PERFORMANCE
requirements, but remains safe.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)