C-2
Name Default
Saved at Power
Failure
Affected by
Defaults
Low (mmHg)
Pediatric 50
Neonate 25
NIBP-Dia High
(mmHg)
Adult 90
Yes Yes
Pediatric 70
Neonate 60
NIBP-Dia Low
(mmHg)
Adult 50
Yes Yes
Pediatric 40
Neonate 20
C.1.2 SpO
2
Name Default Config
Saved at Power
Failure
Affected by
Defaults
Alarm On/Off On
Yes Yes
Alarm Priority Med
SpO
2
High
Adult/pediatric:100
Neonate: 95
SpO
2
Low 90
Desat Low 80
Sat-Seconds (Nellcor) 0s
NIBP Simultaneous Off
Sweep 25 mm/sec
Beat Volume 2
SpO
2
Sensor Off Lev.
Monitor mode: Low
Spot Check mode: Off
Yes No
Sensitivity (Mindray) Med
Yes Yes
Sensitivity (Masimo) Normal
Averaging (Masimo) 8 s