D-3
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause Solution
NIBP Air
Pressure Err
Yes Low
An error occurred to
the air pressure.
Check the air pressure.
Restart the monitor and
retry. If the error remains,
contact your service
personnel.
NIBP Weak
Signal
Yes Low
The patient’s pulse is
weak or the cuff is
loose.
Check the patient’s
condition and change the
cuff application site. If the
error persists, replace the
cuff.
NIBP
Overrange
Yes Low
The measured NIBP
value is not within
the specified range.
Contact your service
personnel.
NIBP
Excessive
Motion
Yes Low
Patient’s arm moves
too much.
Check the patient’s
condition and reduce the
patient motion.
NIBP Cuff
Overpress.
Yes Low
The NIBP airway may
be occluded.
Check the airway and
measure again.
NIBP Signal
Saturated
Yes Low
The NIBP signal is
saturated.
Check the patient’s
condition and reduce the
patient motion or other
sources.
NIBP Air Leak Yes Low
The NIBP airway may
leak air.
The NIBP cuff is not
properly connected, or
there is a leak in the
airway.
NIBP Equip Err Yes High
System error; or
pump, A/D sampling
or pressure
transducer error; or
pointer error during
software running.
Check the patient’s
condition and NIBP
connections, or replace
the cuff.
NIBP Timeout Yes Low
Time is out. In
Adult/Pediatric
mode, the
measurement time
is over 120 seconds;
in neonate mode,
the time is over 90
seconds.