D-5
SpO
2
Comm
Stop
(Mindray,
Masimo,
Nellcor)
No High
An error occurred
to the SpO
2
module, or there is
a problem with
the
communications
between the
module and the
monitor.
Restart the monitor. If the
error remains, contact
your service personnel.
SpO
2
Comm
Abnormal
(Mindray,
Masimo,
Nellcor)
Yes High
SpO
2
Init Err Yes High
SpO
2
Board
Fault
(Masimo)
No Low
There is a problem
with the SpO
2
measurement
board.
Do not use the module
and contact your service
personnel.
SpO
2
Low
Signal
(Masimo)
No Low
The SpO
2
signal is
too low or too
weak.
Adjust the sensor
application site.
SpO
2
Weak
Signal
(Nellcor)
No Low
SpO
2
Interference
(Masimo)
No Low
The SpO
2
signal
has been
interfered.
Check for any possible
sources of signal noise
around the sensor and
check the patient for great
motion.
SpO
2
Comm
Err (Mindray,
Masimo,
Nellcor)
Yes High
An error occurred
to the SpO2
module, or there is
a problem with
the
communications
between the
module and the
monitor.
Restart the monitor. If the
error remains, contact our
service personnel.
SpO
2
Limit Err
(Mindray,
Masimo,
Nellcor)
No Low
The alarm limit of
SpO2 is changed
inadvertently.
Contact your service
personnel.
PR Limit Err
(Mindray,
Masimo,
Nellcor)
No Low
The alarm limit of
PR is changed
inadvertently.
Contact your service
personnel.