D-7
Temp
Prediction
Err
Yes Low
The measuring
operation is improper.
Retry to measure.
Temp
SelfTest Err
No High
An error occurs during
Temp module
initialization.
Replace the module.
Temp Over
High Limit
No High
The patient's
temperature is too
high, or an error
occurs.
Reduce the patient's
temperature, or replace
the module.
Temp Over
Low Limit
No High
The patient's
temperature is too
low, or an error occurs.
Raise the patient's
temperature, or replace
the module.
Temp No
Probe
Yes Med
The probe is
disconnected.
Reconnect the probe.
Temp Probe
Misplaced
Yes Med
The temperature
probe is not well
placed, or not inserted
into probe well.
1 Check if the probe well is
installed.
2 Properly re-insert the
probe into probe well.
Temp
Measuring
Timeout
Yes Med
The measuring time is
over 5 minutes in
Monitor mode.
Return the sensor to the
probe well, and take a
temperature again.
D.2.4 CO
2
Alarm Messages
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause Solution
CO
2
Comm
Err
No High
An error occurred to
the CO
2
module, or
there is a problem
with the
communications
between the module
and the monitor.
Restart the monitor. If the
error remains, contact our
service personnel.
CO
2
Comm
Abnormal
No High
CO
2
Comm
Stop
No High
CO
2
Init Err No High
CO
2
SelfTest
Err
No High An error occurs during
CO2 module
initialization.
Replace the module.