D-9
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause Solution
Rec Head
Wrong Pos.
Yes Low
The thermal head of
the recorder is in
wrong position.
Restore the control lever
of the recorder to its
previous position.
Recorder out
of paper
Yes Low
The recorder paper is
used up.
Replace with a new paper
roll.
D.2.6 Power Alarm Messages
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause Solution
12V Too High No High
There is a
problem with the
system power
supply.
Restart the monitor. If the
problem still exists, contact
your service personnel.
12V Too Low No High
5V Too High No High
5V Too Low No High
3.3V Too High No High
3.3V Too Low No High
Battery Too
Low
No Med
The battery
charge is too low.
Connect the monitor to an AC
power source and allow the
batteries to charge
immediately.
Battery
Depleted
No High
The battery
charge is almost
depleted.
Power Board
Comm Err
No High
No data from
power module
has been received
for 5 seconds.
Restart the monitor. If the
problem still exists, contact
your service personnel.
RT Clock Need
Reset
No Low
Button cell
problem
Reset the system time and
restart the monitor. If the
problem still exists, contact
your service personnel.
PWR
interrupted.
Check meas.
State.
Yes Low Power supply to
the monitor was
interupted.
Check the measurements
when the monitor restarts. If
the problem still exists,
contact your service
personnel or Mindray.