EasyManua.ls Logo

Mindray VS-900 - Page 63

Mindray VS-900
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-7
Parameter Low alarm Limit High alarm Limit Auto-set alarm
limit
Range
Adult/
Pediatric
Neonate Adult/
Pediatric
Neonate
NIBP-D
(Dia×0.68)+
6mmHg
(Dia-15) or
20mmHg,
whichever is
greater
(Dia×0.86)+
32mmHg
(Dia+15) or
80mmHg,
whichever is
smaller
Adult:
25 to 225 mmHg
Pediatric:
25 to 150 mmHg
Neonate:
20 to 90 mmHg
NIBP-M
(Mean×0.68)+
8mmHg
(Mean-15) or
35mmHg,
whichever is
greater
(Mean×0.86)
+35mmHg
(Mean+15) or
95mmHg,
whichever is
smaller
Adult:
30 to 245 mmHg
Pediatric:
30 to 180 mmHg
Neonate:
25 to 105 mmHg
Temp
(T-0.5)°C (T-0.5) °C (T+0.5) °C (T+0.5)°C
Same as the
measurement
range
EtCO
2
0 to 32: remains
the same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement
range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement
range
36 to 45: (EtCO
2
- 6)
36 to 45:
(EtCO
2
- 6)
36 to 45:
(EtCO
2
+ 6)
36 to 45: (EtCO
2
+ 6)
Same as the
measurement
range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement
range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement
range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement
range

Table of Contents