6-7
Parameter Low alarm Limit High alarm Limit Auto-set alarm
limit
Range
Adult/
Pediatric
Neonate Adult/
Pediatric
Neonate
NIBP-D
(Dia×0.68)
6mmHg
(Dia-15) or
20mmHg
whichever is
greater
(Dia×0.86)
32mmHg
(Dia15) or
80mmHg
whichever is
smaller
Adult:
25 to 225 mmHg
Pediatric:
25 to 150 mmHg
Neonate:
20 to 90 mmHg
NIBP-M
(Mean×0.68)
8mmHg
(Mean-15) or
35mmHg
whichever is
greater
(Mean×0.86)
35mmHg
(Mean15) or
95mmHg
whichever is
smaller
Adult:
30 to 245 mmHg
Pediatric:
30 to 180 mmHg
Neonate:
25 to 105 mmHg
Temp
(T0.5)°C (T0.5) °C (T0.5) °C (T0.5)°C
Same as the
measurement
range
EtCO
2
0 to 32: remains
the same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement
range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement
range
36 to 45: (EtCO
2
- 6)
36 to 45:
(EtCO
2
- 6)
36 to 45:
(EtCO
2
+ 6)
36 to 45: (EtCO
2
+ 6)
Same as the
measurement
range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement
range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement
range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement
range