VWALL:24V +/- 5% (only when AC present)
Battery Voltage: 13 21V
Battery Temperature: 27°C / 0 to 60°C
Charging State:
o OFF ------------------------------------- 0
o SOFT_START ------------------------ 1
o TOP_OFF ----------------------------- 2
o RAPID_CHARGE ------------------- 3
o TRICKLE ----------------------------- 4
o MAINTENANCE -------------------- 5
o RECONDITION ---------------------- 6
o COOL_DOWN ----------------------- 7
o DRIVE_DONE ----------------------- 8
o RECONDITION_INIT -------------- 9
Norminal Capacity (LMD): 12960mAH (162)
Battery Charge 12960 nominal
* Below this range would be considered a ‘bad’ (needing
replacement) battery.
Above this range should be considered an invalid value.
Enable Debug Console
Allows (if checked) or disables (if unchecked) access to
the System Console window (ALT-GR / Shift / Z keys) for
entering low-level commands for service. Lock-out
offered due to same key combination used for
international character text entry during normal use.

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals