Figure 19.32-ZS3
11. Once the cables are free, set the Power Supply Module aside and prepare for
installation.
Installation:
1. Reconnect the cables to the replacement Power Supply Module.
2. Tuck cables back into chassis and position Power Supply Module for
installation.
3. Reinstall the eight (8) screws. Ensure that the Power Supply Module is securely
connected.
4. Reattach the AC power cord to the socket at the rear of the Power Supply
Module.
5. Install battery per instructions if one exists.
6. Before powering on the cart, please engage the two circuit breakers as follows:
a. Push “in” the circuit breaker on the rear of the ZPAK
b. Connect AC Cord to cart and outlet
c. Flip the cart circuit breaker “on” by pushing in the “1”.