ZS3 Service Manual Page 261 of 295
4. Remove the lower bolt and screws from the bottom of the gas spring. In this
case, the bottom fixing parts of the gas spring are removed, and the gas spring
can be taken out from the stand column.
Figure 19.53-ZS3
5. Reinstall the gas spring in the reverse steps as shown above. During
installation, adjust down the stud bolt of the gas spring and tighten the screws,
to realize that the gas spring handle can normally control the up and down
function of the system.
Figure 19.54-ZS3
Gas spring & Power Supply Module Re-Installation:
1. Reverse all of the previous steps for installing the replacement gas spring, and
re-installing the power supply module.
System Verification:
Lower bolt and screw
Stud bolt
Screw