ZS3 Service Manual
Page 99 of 295
Name
Description
Port
DICOM communication port, 104 by default. Here, the
port should be consistent with that of the storage
commitment server port.
Maximum
Retries
Reserved feature.
Interval Time(s)
Reserved feature.
Timeout
Value: 5-60s, in increments of 5, and 15 by default. The
connection is disabled if there is no image and
information retrieving in 15 seconds after the user
receiving the information or the image.
Add
Add the DICOM service to the service list.
Cancel
Click to cancel parameter preset.
Update
Select an item in the service list, change the
parameters in the above area, and click [Update] to
update the item in the service list.
Service
list
Delete
Click to delete the selected service in the service list
Verify
Click to verify if the two DICOM application entities are
normally connected.
Default
Set the query/retrieve server as the default.
Select an item in the service list, click [Default] and you
can see Y in the Default column.
Exit
Click to exit the setup screen.
HL7 QUERY SERVICE PRESET
HL7 refers to the medical information exchange protocol on the seventh layer
(application layer) of the Model of the Open System Interconnection published by ISO. It
is established by Health Level Seven International in 1987. HL7 is mainly used for
standardizing the communication between HIS/RIS system and the other devices,
reducing the information interconnection costs among hospitals, and improving the data
usability for hospital information system.
The protocol version that the ultrasound system supports: V2.3, V2.4, V2.5, V2.6.
1. Select [Setup][DICOM][DICOM Service Setting][HL7] to enter the page.
2. Select device in the drop-down list, the available devices are services already set in
the system, enter the information.
Click [Add] to add the service to the Service list;
Select an item in the service list, change the parameters in the above area, and
click [Update] to update the item in the service list; click [Cancel] to cancel
changing.

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals