The device has to be installed in an EMC-compliant manner, see Chapter 4.3.
2.4.2 Supply voltage connection
The supply voltage is 24 V DC ±20%.
Max. power consumption:
- PR 5220/00: 6.5 W
- PR 5220/01: 8.5 W
- PR 5220/04: 8.5 W
- PR 5220/06: 8.5 W
- PR 5220/07: 8.5 W
For connection to 230/115 V AC, an external power supply (e.g., Minebea Intec PR 1624/
00 or Phoenix Mini Power) is required.
2.4.3 Protective conductor connection
The protective conductor is connected via the mounting rail.
2.5 RF interference suppression
The device is designed for operation in the industrial sector and can cause
RF interferences in a residential environment, see Chapter 17.4.3. In this case, the
operator can be required to carry out the appropriate measures.
2.6 Failure and excessive stress
If there is any reason to assume that safe operation of the device is no longer ensured,
shut it down and make sure it cannot be used.