Note:
If the mV/V value of the dead load was calculated, or if it is known from the previous
calibration, the value can be overwritten by pressing [by mV/V].
Error messages, calibration see Chapter 16.5.
7.12.7 Calibrating with weight
Note:
If a linearization was carried out (see Chapter 7.12.11), after selection of the line [Max at]
the following tip is displayed:
Cannot be changed here
while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
1. Press the [by load] softkey.
2. Place the CAL weight on the scale.
3. Enter the weight of the CAL weight.
4. Conrm the entries.
7 Commissioning Transmitter Series PR 5220
EN-81 Minebea Intec