9.2.7.1 Saving conguration and calibration data
9.2.7.2 Loading conguration and calibration data into the device
NOTICE
All data which can be congured in the Setup menu are overwritten!
If the le is loaded into several devices, changing the network settings and the host
name is indispensable!
9.3 Resetting the device to the factory settings
Note:
Reset to the factory settings is possible only, when the CAL switch is open. The IP
address and the Hostname remain unaected.
1. Click on [Backup] to create a backup .g. on the notebook.
2. Click on [Save as].
3. Create and open the required directory e.g. on the notebook.
4. Click on [Save] to save the backup le in the relevant directory.
1. Open the CAL switch; see Chapter 7.1.3.1.
2. Click on [Select File] (depending on the Internet browser).
3. e.g.on the notebook, navigate to the folder where the backup le was saved.
4. Select Backup le.
5. Click on [Restore].
The selected le is loaded into the device.
6. Click on the symbol in the Internet browser to return to the WEB menu.
7. Close the CAL switch.
1. Click on .
2. Click on .
9 Extended functions Transmitter Series PR 5220
EN-181 Minebea Intec