16.2 Weight error status
LED Error 1
Arithmetic
Error 7
(negative)
Error 2
Overload
Error 3
(>36 mV)
Error 6
Sense monitoring
Error 9
Communication
Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz Altern. ash. 1Hz
Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz
Flash. 1Hz Flash. 1Hz Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz Altern. ash. 1Hz
Note:
VNC text, see also Chapter 16.1.
Note:
With all other messages, the upper status indicator LED blinks.
16.3 Error messages for xBPI scales
VNC text Error and possible cause Remote dis-
play
Ext. meas. device error No weight values can be read from scale:
- Scale error
Error 3
Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
No values from scale No communication with xBPI scale:
- Cable break.
- Internal scale error.
- The scale is not connected to the power supply.
Error 9
No weight data The weight value is not displayed:
- Another weighing point was selected.
Press to assign the new weighing point to the device.
Error
Transmitter Series PR 5220 16 Error messages
Minebea Intec EN-220

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec PR 5220 Series and is the answer not in the manual?

Minebea Intec PR 5220 Series Specifications

General IconGeneral
BrandMinebea Intec
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals