VNC text Error and possible cause Remote dis-
play
Scale not ready The scale is not ready for weighing:
- The instrument is in the warm-up phase.
- The instrument is in automatic taring mode.
- The instrument was switched on with the scale loaded.
Switch the device o and on again.
Error <
Wrong serial number Serial number of scale does not match the number set in the de-
vice.
Error?
16.4 Error messages for Pendeo® load cells
VNC text Error and possible cause Remote dis-
play
Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
Incompatible units Incompatible units:
- Incorrect calibration values e.g.due to incorrect calibration
Error 8
No values from scale No communication with Pendeo load cell:
- Cable break.
- Internal scale error.
- The junction box is not connected to the supply voltage.
Error 9
Scale not ready The scale is not ready for weighing:
- At least 1 load cell gives an error status or is defective (no
communication).
Error <
Wrong conguration The number of load cells does not match the conguration. Error =
Wrong serial number Serial number of scale does not match the number set in the de-
vice.
Error?
16 Error messages Transmitter Series PR 5220
EN-221 Minebea Intec