Output if <0: set 0 mA = 0 mA, set <4 mA> =
4 mA, set 20 mA = 20 mA, linear = goes below
4 mA down to the limit (with 420 mA)
Output if >Max Output if >Max: set 0 mA = 0 mA, set 4 mA =
4 mA, set <20> mA = 20 mA, linear = exceeds
20 mA up to the limit
Weight at 0/4 mA Weight value for 0/4 mA output
Weight at 20 mA Weight value for 20 mA output
7.12 Calibrating internal weighing point
7.12.1 General information
Transmitters of the PR 5220 series are not approved for calibration.
The legal-for-trade application of PR 5220 with other device has to be veried.
The calibration data are protected by the CAL switch (see Chapter 7.1.3.1).
During calibration, the instrument must be set to gross weight display (reset tare, if
necessary).
7.12.2 Displaying calibration data
7.12.2.1 Overwrite protection via a CAL switch
When the CAL switch is closed, a tool tip is displayed.
Transmitter Series PR 5220 7 Commissioning
Minebea Intec EN-74