7.12.4 Setting maximum load
The maximum load (Max) determines the maximum weight without dead load of the
weight to be measured and the displayed number of digits behind the decimal point.
Normally, Max is less than the load cell capacity (maximum capacity x number of load
cells).
Permissible values for the maximum load are:
Max weight value from 0.00010 to 999999 in t, kg, g, mg, lb or oz.
Maximum weight value must be an integer multiple of the scale interval (1 d). It may have
up to 6 digits and is entered as a numeric value with or without a decimal point.
6. Press the [Yes] softkey to close the menu without calculation of the test value.
7. Press to exit the calibration for good.
A prompt window opens.
8. Press the [Save] softkey to save changes in calibration data.
Transmitter Series PR 5220 7 Commissioning
Minebea Intec EN-78