Description Data
Potential isolation yes
Cable type Twisted pairs, screened, e.g. patch cable CAT5 Autolink (stra-
ight or crossover)
Cable impedance
150
Cable length to HUB Max. 115 m
Certicate ProBus Nutzerorganisation e.V. for HMS Industrial Net-
works AB
Certicate No: Z10931
Note:
The IP address and the subnet mask are set at - [Fieldbus parameter] (see also
Chapters 7.15.3 and 12.2).
The XML le is stored on the CD (Fieldbus directory of the corresponding instrument)
supplied with the unit. The current le is also available for download via the Internet.
http://www.minebea-intec.com
NOTICE
Slave – master device names
A unique device name must be assigned out of the master. This name is given highest
priority when establishing the communication.
When replacing devices or servicing, please note:
Apart from the IP address, the device name must correspond to the one of the
replacement device. Explicit assignment out of the master is required.
Example:
4.5.4 EtherNet-IP interface
The EtherNet-IP interfaces have the type designation PR 5220/07.
It is a complete EtherNet-IP Adapter (slave) with a standard RJ-45 socket for network
connection.
It contains a powerful TCP/IP and EtherNet-IP interface connection with transfer rates of
10 or 100 Mbit/s.
Transmitter Series PR 5220 4 Device installation
Minebea Intec EN-52