16.6 Error numbers @ "LAST_ERROR"
16.6.1 Weighing point error
Number Display Cause
2 in use Weighing point is in use, e.g., by an application.
6 test active Test is active, no weights.
7 cali active Calibration is active, no weights.
8 no standstill No standstill of the scale.
13 tare is active Tare is active.
15 weight is dimmed Weight is not legal-for-trade (<0 or >max.).
16 weight has error Weight error.
17 scale not ready Scale is not ready.
18 cannot tare below zero Taring below zero is not possible.
102 timeout Only on xBPI scales: Time limit exceeded when sending a com-
mand to the scale.
142 calibration active During calibration, taring and zeroing is not possible.
147 no zeroset Zeroset outside of the zero setting range is not possible.
149 Busy The scale is currently busy with another query.
255 hardware error Weighing point is faulty.
16.6.2 Error in the "EasyFill" application
Number Display Cause
0 No error.
1 fatal error Weight error; weighing point is faulty.
2 Material ID invalid Number <0 or >10 entered.
3 Material name inva-
lid
Material name is invalid.
4 Set point invalid Set point + gross value > scale end value, (B1 mode) or gross value.
5 Preset point invalid Value > set point.
6 OVS invalid Value > set point.
8 Neg. tol. invalid Value > set point.
9 Pos. tol. invalid Value > set point.
10 Sequence number
invalid
Sequence number is invalid.
13 Invalid eldbus com-
mand
Fieldbus action is invalid (e.g., simultaneous start and stop com-
mand).
15 Cannot read from
earom
Error when reading a material entry from the hard drive memo-
ry (EAROM) → hardware error