Message Possible causes
This message is displayed when the dead load entered in mV/V plus
scale interval in mV/V is higher than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed when the measurement signal is negative
when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.
This message is displayed when the dead load entered is 5 mV/V.
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed if the weight on the scale is < the dead load
after input of the weight value.
The maximum capacity is not an integer multiple of the scale interval.
16 Error messages Transmitter Series PR 5220
EN-223 Minebea Intec