Message Possible causes
This message is displayed when the dead load entered in mV/V plus
scale interval in mV/V is higher than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed when the measurement signal is negative
when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.
This message is displayed when the dead load entered is 5 mV/V.
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed if the weight on the scale is < the dead load
after input of the weight value.
The maximum capacity is not an integer multiple of the scale interval.
16 Error messages Transmitter Series PR 5220
EN-223 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec PR 5220 Series and is the answer not in the manual?

Minebea Intec PR 5220 Series Specifications

General IconGeneral
BrandMinebea Intec
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals