EasyManuals Logo

Minebea Intec PR 5220 Series User Manual

Minebea Intec PR 5220 Series
235 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
Description Data
Output: Temperature er-
ror
<100 ppm/K
Output: Zero point error 0.05%
Output: Max. error <0.1%
Output: Load
max. 0500
Output: Short-circuit
protected
yes
Output: Potential isolati-
on
yes
Cable type Screened twisted pair (e.g. LifYCY 2x2x0.20)
Cable length <150 m screened
Analog signal "Current output" Analog signal "Voltage output"
Current is supplied directly via the terminals. The voltage level corresponds to the drop across the
external 500 resistor.
4.4.5 Digital inputs
Three passive opto-decoupled inputs are permanently built into the device. The interface
can be congured by software.
All inputs have a common GND (-).
Transmitter Series PR 5220 4 Device installation
Minebea Intec EN-34

Table of Contents