Note:
For additional connections, refer to the PR 1626/60 instrument manual.
If the solder bridge MX8 in the PR 1626/60 is closed (7.5 V DC), the supply at -
[Weighingpoint] - [WP A] - [Calib] - [Param] - [External supply] has to be set to "below
8 V DC." The sense voltage detection is now switched.
NOTICE
Installation in explosion-hazarded areas
The screen of the load cell cable and the screen of the connecting cable must not be
connected inside the junction box whenever regulations governing installation in
explosion-hazarded areas prohibit any connection of the two ends.
Transmitter Series PR 5220 4 Device installation
Minebea Intec EN-44