— W&M Selection legal-for-trade mode: <none>, OIML
(impossible when [Range mode] "Multi-
interval" has been selected or Max has more
than 3 decimals), NSC, NTEP
Note: Transmitters of the PR 5220 series are
not approved for calibration.
— Standstill time
Input of the standstill period: 0.01 s<0.50
s>2.0 s (The range depends on the measuring
time.)
— Standstill range
Input of the standstill range: 0.00 d<1.00
d>10.00 d
— Tare timeout Input of timeout when there is no standstill: 0.1
s<2.5 s>25 s
— Zeroset range ± Range of zero point when there is no
standstill. Input: 0.00 d<50.00 d>10000.00
d
— Zerotrack range
Input of the zerotrack range: 0.00 d<0.25
d>10000.00 d
— Zerotrack step
Input of the zerotrack step: 0.00 d<0.25
d>10.00 d
— Zerotrack time
Input zerotrack time: <0.0 s>25 s
— Overload Input of the weighing range above the
maximum load (Max) without error message:
0999999 d
— Minimum weight
Input of the minimum load: 0 d<50
d>999999 d
— Range mode Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.12.15.1 and 7.12.15.2.
— Range limit 1 Input of the limit 1: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— Range limit 2 Input of the limit 2: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— View (when CAL-switch is closed)
— Scale interval Display only
— Dead load at Display only
— Calibrated at Display only
— Sensitivity (µV/d) Display only
— Param Items as for [Param] (display only)
— Setup Setup of the xBPI scale
— Calibration Calibration of xBPI scale
— Dead load Dead load/preload
— Set Set dead load/preload: Accept, ResError, Abort
— Delete Clear dead load/preload: Accept, ResError,
Abort
7.11.5.2 "xBPI scale" weighing point
Transmitter Series PR 5220 7 Commissioning
Minebea Intec EN-68