PMM 9030/9060/9180 Frequency Extension 11-7
Noise level (Preselector ON)
(Preamplifier OFF)
(Preamplifier ON)
30 to 300 MHz < 5 dBuV (AV)
(120 kHz BW)
300 to 1000 MHz < 7 dBuV (AV)
(120 kHz BW)
1000 to 4000 MHz < 10 dBuV (AV)
(120 kHz BW)
4000 to 6000 MHz < 14 dBuV (AV)
(120 kHz BW)
30 to 300 MHz < -11 dBuV (AV)
(120 kHz BW)
300 to 1000 MHz < -6 dBuV (AV)
(120 kHz BW)
1000 to 4000 MHz < -5 dBuV (AV)
(120 kHz BW)
4000 to 6000 MHz < -1 dBuV (AV)
(120 kHz BW)
Spurious response
< 10 dBuV, < 15 dBuV above 2 GHz,
Measuring Detectors
Peak, Quasi-peak, Average, RMS, C-Avg, C-RMS,
Smart Detector
Level meas. time
Peak, Quasi-peak, RMS and Average
(simultaneous detectors)
1 ms to 30 sec. (CISPR 16-1-1 default)
Measurement accuracy
S/N > 20 dB
30 to 1000 MHz ± 1.0 dB
1 to 3 GHz ± 1.5 dB
3 to 6 GHz ± 2.0 dB
I/O Interface Out
High Speed Optical Link;
RS232 (service only)
Operating temperature
0° to 40°C
Power supply
10 - 15 Volt DC, 2,5A; Li-Ion rechargeable and
interchangeable battery (4 h operations, average)
Dimensions
235 x 105 x 105 mm
Weight
2.2 kg