10-4 Click Mode Operating Instructions
10.3 Enter the Click
Mode
To enter Click Mode, press the corresponding key in the main menu, and
immediately the first window opens as following:
With PMM 9010/03P /30P /60P press the Cond..ed A-B key in the initial
menu to enter the main menu where the Click Mode button appears.
Fig. 10-1 Click
This is the IDLE function, which allows a preliminary, free running
evaluation, at the selected frequency and level.
The Click mode function has four commands:
Start
Clear: cancel and reset all data to start an idle test;
Report
SETUP
Always use Esc button to return to the previous view/condition.
To increase test productivity the PMM 9010, PMM 9010/03P/30P/60P or
PMM 9010F has some very unique features: it allows skipping as many as
2 steps, as it may predict how many clicks would overcome the Lq limit and
may also advice the User if the next step would fail.
Moreover, it makes a true Qpeak measure and not a mere threshold
comparison, as only a fully digital receiver can detect whether a Qpeak
level came from an unsaturated signal or not; an analogue analyzer only
has the final weighted Qpeak.