EasyManuals Logo

NARDA PMM 9010 Series User Manual

NARDA PMM 9010 Series
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
10-14 Click Mode Operating Instructions
10.7.6 Terminate on
To define the automatic stop condition, it is possible to select between two
options: 40 Clicks/Time and Time/Manually.
In the first case, the stop of the step either happens after 40 clicks or after
the defined time has elapsed; in the second case, the step is not
terminated when reaching 40 clicks (thus allowing to test EUTs that are
commanded by a program) but only when the defined time has elapsed. It
is always possible to stop the step, and even the test, manually.
Press Change to flip from one to the other option.
10.7.7 Line
When using a PMM LISN, it’s possible to select the Line on which to
perform the test through the Setup Menu, just selecting “LINE” and
accessing the related sub-menu, as per the following:
In case a 3rd party LISN would be used, just leave “NONE” in the menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NARDA PMM 9010 Series and is the answer not in the manual?

NARDA PMM 9010 Series Specifications

General IconGeneral
BrandNARDA
ModelPMM 9010 Series
CategoryReceiver
LanguageEnglish

Related product manuals