9-8 General Information
9.10 PMM 9010/30P
Main Specifications
Table 9-2 lists the PMM 9010/30P performance specifications.
The following conditions apply to all specifications:
The ambient temperature shall be 0°C to 40°C
For what is concerning the specifications of the full compliance A
and B bands section refer to the specification of the 9010 in the
chapter 1.
TABLE 9-2 Main Specifications for CDE bands
Frequency range
Resolution
Frequency accuracy
30 MHz to 3 GHz
100 Hz
< 2 ppm
RF input
VSWR
10 dB RF att.
0 dB RF att.
Attenuator
Z
in
50 Ω, N fem.
< 1.2, <2.0 over 1 GHz
< 1.2, <2.0 over 1 GHz
0 dB to 50 dB (2dB steps)
Max input level
(without equipment damage)
Sinewave AC voltage
Pulse spectral density
PRF (CISPR 16-1-1)
137 dBµV (1 W)*
97 dBµV/MHz
10Hz
IF bandwidth
Standard 6 dB
CISPR 16-1-1 bandwidth (6 dB)
MIL-STD-461 (option)
3, 10, 30, 100, 300 kHz, 1 MHz
120 kHz
100 kHz, 1 MHz
Noise level
30 MHz to 3 GHz < 8 dBµV (QP)
(120 kHz RBW) < 4 dBµV (AV)
Spurious response
< 15 dBµV
Measurement accuracy
(S/N > 20 dB)
30 MHz to 1 GHz ± 1,0 dB
1 GHz to 3 GHz ± 1,5 dB
Operating temperature
Power supply
0° to 40°C
10 - 15 Volt DC, 2,5A; Optional Li-Ion interchangeable
battery (8 h operations, typical)
Dimensions
235 x 105 x 335 mm
Weight
4.1 kg
(*) with MIN ATT 10 dB