4.1 Using Limiter, Setting Limiter
DP Series/DP Series Type R
81
Operation procedure
1.
Push [Limiter] soft-key to select 2:Irms Limiter (or push I key). Opens the RMS current limiter
set window.
2.
Set the value of RMS current limiter in [Io] item.
3.
Set the [Output OFF] item as [Enable] or [Disable]. When setting to [Enable], also set the time
from the start of limiter operation to the output off.
4.
Close the window.