10. Specifications
PROGRAMMABLE AC POWER SOURCE
274
*13: In the case of 10 V to 150 V / 20 V to 300 V, sine wave, no load, 45 Hz to 65 Hz, 23 °C5 °C.
*14: If the output voltage is higher than the rated value, this is limited (lowered) to satisfy the power
capacity. In the case of 40 Hz or lower or 400 Hz or higher, and that the ambient temperature is
40 °C or higher, the maximum current may decrease.
*15: For the capacitor input type rectified load (crest factor=4), the rated output voltage, and 45 Hz to
65 Hz.
*16: In the case rated output voltage, 50 Hz or 60 Hz, power factor 1. It may reduce the reverse
power flow if ambient temperature is 40 °C or higher. If the output voltage is higher than the
rated value, power capacity is limited to 40 % (continuous, 30 °C to 40 °C), 50 % (continuous,
lower than 30 °C), and 100 % (reverse power flow time 180 s, reverse power flow duty 0.2,
40 °C or lower). About duty, see Figure 10-1. Only for DP series Type R, the reverse power flow
is supported.
*17: In the case that the power input is single-phase 170 V or lower, models with 6 kVA or higher
have the limit on the power capacity in the power running.
*18: For 45 Hz to 65 Hz, the rated output voltage, no load and the resistance load for the maximum
current, and the operating temperature.
*19: For 40 Hz to 550 Hz, sine wave, the rated output voltage, the resistance load for the maximum
current at 55 Hz, and 55 Hz reference.
*20: Set for the L1 phase. The component of the phase angle setting is added for the other phases.
*21: In the case of 50 V or higher, sine wave, and same load condition and voltage setting for all
phases.
*22: In the case of the AC mode and 23 °C5 °C.