EasyManua.ls Logo

NT-MDT Solver Next - Selecting Lithography Mode

NT-MDT Solver Next
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Solver NEXT SPM. Instruction Manual
174
8.2.2.1. Selecting Lithography Mode
To select the lithography mode, perform the following steps:
1. In the controller configuration drop-down list of the Nova P9 program, select Contact
(Fig. 8-10).
Fig. 8-10. Selecting the controller configuration
This will result in automatic configuring the software to operate in contact modes. The
parameters will be adjusted as follows:
Output of the feedback loop – DFL signal;
SetPoint parameter (proportional to the force applied to the sample) – 2 nA;
Feedback Gain – minus 10.
2. Adjust the SetPoint level for the contact mode as follows:
a. Break the feedback loop (the button will change its appearance to Off);
b. Adjust the SetPoint level to be 0.5 lower than the current DFL level;
c. Close the feedback loop (the button will change its appearance to On);
d. Increase the SetPoint level and watch extension of the Z scanner section in the
indicator of the Main Operations panel (Fig. 8-11). Set the
SetPoint parameter at
the level when extension of the Z scanner section stops.
Fig. 8-11. Scanner extension indicator
3. Adjust the working level of the feedback gain (Gain parameter) as follows:
a. Double-click in the input field of the Gain parameter. This will open the adjusting
slider of the feedback gain (Fig. 8-12).
Fig. 8-12. Adjusting slider of the
Gain parameter
b. Decrease the Gain level gradually and watch the input of the feedback loop in the
sofware oscilloscope.

Table of Contents