Chapter 3. Operating Conditions
13
3. Operating Conditions
The SPM should be installed in a room of total area larger 6 sq. meters with range of purity
better 8 (by ISO 14644-1-2002).
To provide for the qualitative and safe operation of the instrument, it is recommended to
observe the following conditions:
environment temperature:
ultimate range (15–35) °С;
recommended range (20–25) °С.
temperature drift: less 1˚C per hour;
relative humidity:
ultimate range (20–80) % (moisture condensation prevented);
recommended range (30–50) % (moisture condensation prevented).
vibration amplitude
in the band 1 ÷ 1000 Hz less 0.5 µm;
atmospheric pressure (760±25) mm Hg/(100±4) kPa;
the work area should be supplied with a protective grounding circuitry and with
grounded electric mains (110/220 V);
electric mains:
voltage (110±10)/(220±22) V;
frequency (50.0±0.4)/(60±0.5) Hz;
the room should be protected from mechanical vibrations and acoustic noises, either
internal or external;
vibration criterion VC-C, 12.5 µm/s (one-third octave band criterion);