Solver NEXT SPM. Instruction Manual
202
Fig. 8-59. Image of the lateral force distribution
8.3.2.2. Selecting the lithography method
To select the lithography method, perform the following steps:
1. In the controller configuration drop-down list of the Nova P9 program, select
SemiContact (Fig. 8-60).
Fig. 8-60. Selecting controller configuration
This will result in automatic configuring the software to operate in semicontact modes. The
parameters will be adjusted as follows:
● Input of the feedback loop – Mag signal;
● SetPoint parameter (proportional to the force applied to the sample) – 10 nA;
● Feedback Gain – minus 4.
2. Open the Litho window by clicking the button in the Main Operations panel
(Fig. 8-61).