1020 Indicator
4.2. First use of indicator -Calibration-
Calibration settings are used to check, delete and set calibration points.
First, enter the Indicator Setup menu as decribed on page 14, and press the DOWN
key to go to Calibration and press Enter. If you are already in the Indicator Setup
menu, use the DOWN key to go to Calibration and press Enter.
Enter CAL code (number) by using the UP key and confirm with Enter.
CAL code shows on the lower right of the screen. Every time calibration settings are
changed, the CAL code automatically levels up by 1.
Example CAL:0
CAL Code
20