5.1.2. Menu Settings -System Setup - Indicator Setup - continue-
1020 Indicator
Dual Point
A calibration with 2 points.
See page 11,12 and 13 for calibration instructions.
Multi Point
A calibration with more than 2 point. Calibration upto 10 points is possible. Multi
point calibration is mostly used if the weigher signal is not linear. Also excisting
points can be deleted or replaced in this menu.
Multipoint calibration instructions
Before you proceed, make sure the weigher is unloaded.
Deleting an existing calibration point: