EasyManua.ls Logo

PENKO 1020 - Calibrate the Zero Point; Calibration Second or Higher Calibration Point

PENKO 1020
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.1.2. Menu Settings -System Setup - Indicator Setup - continue-
1020 Indicator
Calibrate the zero point:
Make sure the weigher is unloaded and press Enter.
Calibration second or higher calibration point:
An actual reference value is needed. For this example, a reference value of 1020
kg was used.
Use the UP, DOWN and LEFT key to enter the reference value. The UP and
DOWN keys are used for changing the number (1-9), the LEFT key is used for
changing the position of the cursor.
Load the weigher with the reference value and press the Enter key.
The calibration was successful when the following screen is visible:
47

Table of Contents

Related product manuals