LIFEPAK 500 automated external defibrillator Replacement Procedures
11-50
Removing Biphasic PCB—Biphasic AED Page 2 of 2
2. Disconnect the seven spade-lugs P101, P102, P103, P104, P105, P107, and
P108, from J101, J102, J103, J104, J105, J107, and J108, respectively, on
the underside of the Biphasic PCB. Do not pull on the wires.
3. Disconnect P106 of the Data Cable from J106 of the Biphasic PCB.
4. Carefully remove the Biphasic PCB and place it in a static-protective bag.
5. After removing the Biphasic PCB, jump to the appropriate replacement
procedure:
– Replacing Inductive Resistor—Biphasic AED
– Removing Energy Storage Capacitor—Biphasic AED
– Installing Biphasic PCB—Biphasic AED