LIFEPAK 500 automated external defibrillator Replacement Procedures
11-59
Removing Energy Storage Capacitor—Monophasic AED Page 3 of 3
3. Install two new capacitor tie-wraps and tightly cinch the tie-wraps.
After installing the new Energy Storage Capacitor, return to the Installing Main
PCB—Monophasic AED procedure and then to the Reassembling Case
procedure.