LIFEPAK 500 automated external defibrillator Troubleshooting
8-15
Fault Code Table Page 12 of 20
Code Fault Description
Corrective
Action
Code
5908 safmon_cont_mon_task_did_not_run 10
5909 safmon_bad_rtc 9, 2
590a safemon_defib_cap_unexpected_charge 2
590b safemon_excessive_battery_current 10, 2
590f zpredicted_out_of_spec 6, 2
5a00 data_com_fault_minor 10
5a01 data_com_mem_mgmt_invalid_buffer_ptr 10
5a02 al_locked_large_buffer_discarded 10
5c00 util_fault_minor 10
5f00 bte_fault_minor 6
5f01 bte_hv_mon0_mismatch 10, 11, 2
5f02 bte_failed_xfer_input 10, 11, 2
5f03 bte_failed_dump_input 10, 11, 2
5f04 bte_failed_htest_input 10, 11, 2
5f05 bte_xfer_vcap_invalid 10, 11
5f06 bte_phase1_imped_low 10, 11
5f07 bte_phase1_imped_high 10, 11
5f08 bte_phase2_imped_low 10, 11
5f09 bte_phase2_imped_high 10, 11
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB