2-18
VIBXPERT II 05.2012
Getting started - Basic settings
Left:
Basic settings for the route/tem-
plate operating mode
Right:
Options for saving
Standard vibration sensor
Hereyoudenethesensortobeusedforallvibrationmeasurements
on a route. The settings in OMNITREND are ignored. The 'Triax sen-
sor' option is disabled.
Standardsensorisactivewhen,inthetaskselectionscreen
- Channel A / Channel B is marked red,
-Sensoriconatthetopedgeofthescreenlightsuporange.
Use the standard sensor for measuring or deactivate this option.
Otherwise, incorrect measurements may be the result.
Save settings
• Clickon'Saving'tosetupthefollowingoptions:
AutoSave
This shortens the duration of the entire measurement.
-
enaBled: The selected time (0 ... 10s) corresponds to the waiting
time until the result is automatically saved.
-
disaBled: The result must be manually saved.
Stop AutoSave on...
- alarm, warning, prewarning: When the limit value is exceeded, the
AutoSavemodeisdeactivated.
-
never: The result is always stored automatically.
Toacceptthesettingsinthe'RouteSetup'andtoclosethemenu,press
MENU and click on 'OK'.