2-28
VIBXPERT II 05.2012
Analog out/headphones
At the analog output (yellow connector), the vibration signal can be
picked up with a suitable analysis device (oscilloscope) or listened to
viaheadphones(e.g.VIB6.670).Theoutputisalwaysthepure,non-
integrated signal without the DC component.
To use the yellow connector as the analog output, it must be enabled
andparameterizedintheDeviceSetup:
• Clickon'AnalogOut/Headphone'intheDeviceSetup.
• Enabletheanalogoutput.
The signal will be output at the analog output until the measure-
ment screen is closed.
• Ifneeded,adjustthestandardnormalizationfortheaccelerometer.
Youcanchoosefrom:10mV/g,100mV/g,500mV/g*,1mV/ms
-
²
Amaximumoutputvoltageof ±4Vprovides themaximumpos-
sible measurement range, for example of:
80m/s
2
at500mV/g
400m/s²at100mV/g
• Ifthevibrationsignalismeasuredwithadifferentsensortype,set
a suitable gain factor in the 'Default amplification' menu (x0.1 / x1
/x5/x10).
To connect the analysis device, use the cable for the analog output
(VIB5.431):
* Only for sensors with a
sensitivity of 5.35 µA/ms
-2
Normalization and amplification for
Analog Out
Getting started - Basic settings
Headphones - VIB 6.670
VIB 5.431
BNC